Filter series FT

®

FT-3G-H2

FT-3SS-H2

FT-0,1GF-H2

Part number

01 F 3006(a)

01 F 3008(a)

01 F 3007(a)

Filter finess/material of filter element

3µm / glass

3µm / stainl. steel

0,1µm / glassfibre

Power supply

230 - 240V 50Hz, 350VA or 115V 60Hz (a)

Electrical connection

2,5mm

2

terminals, 2x PG13 cable glands

Protection classification/electrical
standard

IP44-EN60529 / EN61010, EN60519-1

Ambient temperature max.

+50°C

Sample temperature max.

+180°C

Ready for operation

after approx. 30 minutes

Temperature alarm contact

alarm point

T 30°C at T

set

, contact rating 250V, 3A, 0,25A=

Regulation of temperature

thermostat (PT100 as option)

Sample connections

G1/4” i

(DIN ISO 228/1)

Stagnant space

65cm³

Filter surface

70cm³

Ma. pressure

4 bar g

Dimensions (w x h x d) / weight

350 x 300 x 140mm / 6kg

Method of mounting

wall mounting

Material of sample contacting parts

PTFE, glass, FPM

(a) 115V-Version