RECEPTION AND STORAGE
The universal filter series FT-
…-H2 is normally delivered in one packaging unit:
The universal filter series FT-
…-H2 should be removed carefully from the packaging and checked
immediately for completeness against the delivery note.
Check the goods for any damage incurred during transport and if necessary inform your transport
insurer of any damage.
10
PREPARATION AND INSTALLATION
Locate the universal filter series FT-
…-H2 in such a way that there is adequate space for removing the
cover and replacing the filter elements. Fix the aluminium plate with 4 screws.
Make certain that the universal filter series FT-
…-H2 is easily accessible so that you can carry out any
subsequent maintenance work without trouble.
11
MOUNTING
For mounting the filter we recommend the following procedure:
Remove the cover of the heated filter after opening the two clamping devices.
Disconnect the thermal conductivity jaws on the left and right side of the filter head in by
loosening the screws.
In order
to connect the sample lines, screw in suitably sized threaded connectors with a G ¼”
connecting thread using PTFE sealing tape.
N O T E !
Use only connectors approved for 180°C!
Make sure that the connection is leak proof!
Factory set the filter is mounted with inlet on the right and outlet on the left side. To turn that by 180 °
consider the following:
Loosen, the 2 screws (M4), placed at the outside of the the top plate.
Pull out the filter from heating jacket
Loosen the 2 screws, at the inside of the top plate
Turn the filter by 180° and fix it with the inner screws at the top plate
Put in the filter into the aluminium heating jacket and fix it with the two outer screws at the top
plate
Connect the sample lines and assemble the two thermal conductivity jaws on the in and outlet
once again.
Put the cover of the heated filter on top again and close the clamping devices.