beautypg.com

Electrical connection – M&C TechGroup FT-...-H2 Operator's manual User Manual

Page 11

background image

11

Gas sampling and gas conditioning technology

5-2.1.1-ME

12

ELECTRICAL CONNECTION

W A R N I N G !

When connecting the equipment, please ensure that the supply
voltage is identical with the information provided on the model type
plate.

After mounting the filter, the posssibility of getting in touch with
live parts has to be prevented!

The use of temperature resistant cable is necessary !

N O T E !

Attention must be paid to the requirements of IEC 364 (DIN VDE
0100) when setting high-power electrical units with nominal
voltages of up to 1000 V, together with the associated standards
and stipulations.

A main switch and matching fuse must be provided externally!

The main circuit must be equipped with a fuse corresponding to
the nominal current (over current protection), for electrical details
see technical data.

Remove the lid of the electrical connection box. The electrical connection layout is located in

the lid.

Insert the mains cable (min. 3 x 1.5 mm²) through the cable gland and connect to the

appropriate terminals (Fig.2).

Insert the signal cable through the cable gland and connect to the appropriate terminals.

Screw lid back on.