Technical data – M&C TechGroup ECP20-2 Operator's manual User Manual
Page 9
![background image](/manuals/255967/9/background.png)
9
Gas sampling and gas conditioning technology
3-1.10-ME
8
TECHNICAL DATA
Electric Gas Cooler ECP
Type ECP20-1
Type ECP20-2
Sample outlet dew point
range of adjustment: +2 °C ..... +15 °C, factory setting: +5 °C
Dew point stability
at const. conditions: < ±0,1°C
Sample inlet temperature
**max. 180°C
Sample inlet dew point
**max. 80°C
Gas flow rate
**max.250l/h
**max.500l/h
Number of heat exchangers
1
2
Material of heat exchangers
Duran glass or PVDF or stainless steel 316
Ambient temperature
**+5°C up to +40°C
Stockage temperature
-20 to +60°C
Admissible gas pressure
max. 3bar g. with glass and PVDF
max. 10bar with stainless steel
Total cooling capacity at 25°C a.t.
50KJ/h
80KJ/h
Dead volume/heat exchanger
approx. 80cm
3
Sample gas connection
glass: for tube Ø6mm, PVDF/SS316: G1/4“i
Condensate connection
glass: for tube Ø12mm, PVD
F/SS316: G3/8“i
Ready for operation
< 20 minutes
Mains power supply
switch 230V 50Hz or 115V 60Hz,(±10%)
Power consumption
150VA
250VA
Electrical connection
terminal 2,5mm
2
, cable glands 2 X PG11
Status alarm
contact rating 250V, 2A, 500VA, 50W
alarm point
3°C to T
soll
Electrical protection
Fuse 2 x 4A
T
Case protection
IP 54 (EN 60529)
Method of mounting
surface mounting
Casing dimensions (HxWxD)
210mm X 385mm X 226mm
Wight
10Kg
12,5Kg
Electrical equipment standard
EN 61010
* Standard,
** Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C
ambient temperature and an outlet dew point of 5 °C.