beautypg.com

Technical data – M&C TechGroup ECP20-2 Operator's manual User Manual

Page 9

background image

9

Gas sampling and gas conditioning technology

3-1.10-ME

8

TECHNICAL DATA

Electric Gas Cooler ECP

Type ECP20-1

Type ECP20-2

Sample outlet dew point

range of adjustment: +2 °C ..... +15 °C, factory setting: +5 °C

Dew point stability

at const. conditions: < ±0,1°C

Sample inlet temperature

**max. 180°C

Sample inlet dew point

**max. 80°C

Gas flow rate

**max.250l/h

**max.500l/h

Number of heat exchangers

1

2

Material of heat exchangers

Duran glass or PVDF or stainless steel 316

Ambient temperature

**+5°C up to +40°C

Stockage temperature

-20 to +60°C

Admissible gas pressure

max. 3bar g. with glass and PVDF

max. 10bar with stainless steel

Total cooling capacity at 25°C a.t.

50KJ/h

80KJ/h

Dead volume/heat exchanger

approx. 80cm

3

Sample gas connection

glass: for tube Ø6mm, PVDF/SS316: G1/4“i

Condensate connection

glass: for tube Ø12mm, PVD

F/SS316: G3/8“i

Ready for operation

< 20 minutes

Mains power supply

switch 230V 50Hz or 115V 60Hz,(±10%)

Power consumption

150VA

250VA

Electrical connection

terminal 2,5mm

2

, cable glands 2 X PG11

Status alarm

contact rating 250V, 2A, 500VA, 50W

alarm point

3°C to T

soll

Electrical protection

Fuse 2 x 4A

T

Case protection

IP 54 (EN 60529)

Method of mounting

surface mounting

Casing dimensions (HxWxD)

210mm X 385mm X 226mm

Wight

10Kg

12,5Kg

Electrical equipment standard

EN 61010


* Standard,
** Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C

ambient temperature and an outlet dew point of 5 °C.