Closing down, Maintenance – M&C TechGroup ECP20-2 Operator's manual User Manual
Page 16

16
Gas sampling and gas conditioning technology
3-1.10-ME
14
CLOSING DOWN
N O T E !
The area in which the cooler is situated when not in use must be
kept free of frost at all times!
If the cooler unit is putting out of action for a short time no particular measures need to be taken.
We recommended sweeping the cooler with inert gas or ambient air while the unit is putting out of
action for a longer time. Condensate has to be removed completely from the cooler.
W A R N I N G !
Aggressive condensate possible.
Wear protective glasses and proper protective clothing!
15
MAINTENANCE
Before the maintenance work is carried out, it is necessary that the specific safety procedures
pertaining to the system and operational process be observed!
W A R N I N G !
It is necessary to take the ECP electric gas cooler off the mains
before any assembly, maintenance or repair work is carried out!
The ECP 20 Gas Cooler requires no particular routine maintenance. Depending on the quality of the
ambient air the cooling fin block should be blown out with compressed air from time to time.