beautypg.com

Technical data – M&C TechGroup EC-EX Operator's manual User Manual

Page 9

background image

9

Gas sampling and gas conditioning technology

3-4.1ME EC-EX 9904

10

TECHNICAL DATA

Sample outlet dew point

range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C

Dew point stability

at constant conditions

0,25°C

Sample inlet temperature

**Max. + 180°C

Sample inlet dew point

**Max. +80°C

Gas flow rate per heat exchanger

**Max. 250l/h

Number of heat exchangers

1*,

installation of max. 4 heat exchangers possible

Material of heat exchangers

Borosilicate glass or PVDF or stainless steel 316

Ambient temperature

**+0 to +45°C

Admissible gas pressure

with glass and PVDF: max. 3 bar
with stainless steel

: 10 bar*

with option EC-FD : 1,2 bar

Total cooling power

Max. 520 KJ/h at +25°C

Dead volume per heat exchanger

70 ml

Sample gas connection glass:

PVDF:


stainless steel:

For tube

6mm*,

option: 8mm, 10mm or hose connector

G 1/4“i,

option: hose connector

G 1/4“i*,

option: NPT or hose connector

Condensate connection

glass:


PVDF:


stainless steel:

For tube

12mm*,

option: 8mm, 10mm or hose connector

G 3/8“i,

option: hose connector

G 3/8“i*,

option: NPT or hose connector

Ready for operation

< 30 minutes

Power consumption

280VA, start up current at 230V = 8,1A at 115V= 17A

Mains power supply

230V 50Hz +/- 10% *, or 115V 50-60Hz +/- 10%

Electrical connections

Terminal 2,5 mm²

Status alarm

1 changeover contact, max. 230V 2A AC/DC 100VA, 50W

Ex-Protection

II 2G EEx p e d q [ib] IIC T4

Casing colour

RAL 9003

Method of mounting

19“ rack or wall mounting

Casing dimensions

84 TE x 7 HE x 450mm, 40 kg

Refrigerant

R134A 100% CFC free

Electrical equipment standard

EN 61010


* standard

Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C
ambient temperature and an outlet dew point of 5 °C.