Technical data – M&C TechGroup EC-EX Operator's manual User Manual
Page 9
9
Gas sampling and gas conditioning technology
3-4.1ME EC-EX 9904
10
TECHNICAL DATA
Sample outlet dew point
range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C
Dew point stability
at constant conditions
0,25°C
Sample inlet temperature
**Max. + 180°C
Sample inlet dew point
**Max. +80°C
Gas flow rate per heat exchanger
**Max. 250l/h
Number of heat exchangers
1*,
installation of max. 4 heat exchangers possible
Material of heat exchangers
Borosilicate glass or PVDF or stainless steel 316
Ambient temperature
**+0 to +45°C
Admissible gas pressure
with glass and PVDF: max. 3 bar
with stainless steel
: 10 bar*
with option EC-FD : 1,2 bar
Total cooling power
Max. 520 KJ/h at +25°C
Dead volume per heat exchanger
70 ml
Sample gas connection glass:
PVDF:
stainless steel:
For tube
6mm*,
option: 8mm, 10mm or hose connector
G 1/4“i,
option: hose connector
G 1/4“i*,
option: NPT or hose connector
Condensate connection
glass:
PVDF:
stainless steel:
For tube
12mm*,
option: 8mm, 10mm or hose connector
G 3/8“i,
option: hose connector
G 3/8“i*,
option: NPT or hose connector
Ready for operation
< 30 minutes
Power consumption
280VA, start up current at 230V = 8,1A at 115V= 17A
Mains power supply
230V 50Hz +/- 10% *, or 115V 50-60Hz +/- 10%
Electrical connections
Terminal 2,5 mm²
Status alarm
1 changeover contact, max. 230V 2A AC/DC 100VA, 50W
Ex-Protection
II 2G EEx p e d q [ib] IIC T4
Casing colour
RAL 9003
Method of mounting
19“ rack or wall mounting
Casing dimensions
84 TE x 7 HE x 450mm, 40 kg
Refrigerant
R134A 100% CFC free
Electrical equipment standard
EN 61010
* standard
Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C
ambient temperature and an outlet dew point of 5 °C.