Gas sampling and gas conditioning technology

3-4.1ME EC-EX 9904

13

RECEPTION AND STORAGE

The EC-EX gas cooler is a complete pre-installed unit.

Carefully inspect the cooler and any special accessories included with it immediately on arrival by

removing them from the packing and checking for missing articles against the packing list!

Check the items for any damage in transit and, if required, inform the shipping insurance company

immediately of the damage found!

N O T E !

The cooler must be stored in a weatherproof frost-free area!

N O T E !

During transport and when in storage, the cooler must always be stood up
with the transport feet positioned underneath so that the oil in the closed
compressor circuit cannot run out of this compressor case. If the cooler is
transported on its back by mistake, it must stood in the operating position
for approx. 24 hours before it is switched on!



14

INSTALLATION INSTRUCTIONS

The EC-EX cooler is equally suitable for wall mounting or for installation in a 19" rack.

N O T E !

The compressor cooler EC-Ex is suitable for using in hazardous Group II
Category 2G. Attention must be paid to the certificate of conformity KEMA
03ATEX2113 (see appendix).

N O T E !


Please state the desired type of mounting when ordering so that the LED
function display can be positioned to match at the factory!

N O T E !

The operating position for this cooler is exclusively vertical. This is the
only way to ensure proper separation and removal of condensate in the
heat exchangers. During transport and installation, the cooler must
always be stood up with the transport feet positioned underneath so that
the oil in the closed compressor circuit cannot run out of the compressor
case.

The cooler should be kept away from sources of heat and well ventilated
when installed, so that condensation from warmth will not occur and
interfere with operation.

The minimum installation dimensions (fig. 2) must be followed without
fail. If the unit is installed in the open, the cooler must be installed in a
housing that is frost-free in winter and adequately ventilated In summer.
avoid locating the unit in direct sunlight.

Unheated gas sample lines must be provided with slope up to the cooler.
In that case pre-separation of the condensate is not required.

Connect the heated sample line with sufficient thermal decoupling to the
cooler!