SUPPLY CONNECTIONS

15.1

HOSE CONNECTIONS

The hoses for the heat exchangers are laid out as per Figure 3.

N O T E !

Do not mix up the hose connections; the inlet and outlet connections of the
heat exchangers are marked with arrows;
Ensure that the connections are sealed adequately;
To ensure free removal of the condensate, ensure that the listed diameters
for the condensate removal lines are not reduced!


The connection for sample gas inlet and outlet happens on the upper part of the heat exchangers. For
possible connectors see technical data (10.).

Correspondingly tube or hose connectors are optional available by M&C.
Ensure that the connections are sealed adequately by noting the following:

Borosilicate glass heat exchangers with connections GL 18-6 respectively GL 25-12

Before assembly, check the GL coupling rings to see if the PTFE/silicon locking rings have been

damaged.

The sealing rings should be installed with the PTFE side facing the medium.

To grant a functional and unproblematic mounting we recommend to use union pieces with taper

pipe thread type R according to DIN 2999/1 in connection with suitable sealing tape.

PVDF respectively stainless steel heat exchangers with G 1/4“i respectively G 3/8“i

The correspondingly dimensioned tube respectively hose couplings with threaded connections

have to be screwed in with PTFE thread sealing tape.

N O T E !


When fixing the connectors in the PVDF heat exchanger hold up with a
wrench at the pane of the bolt head!


Option: stainless steel heat exchanger with NPT

The heat exchangers with NPT threaded connectors are marked with circulated notches.

The NPT thread must be screwed in with sealant or fixed with adhesive.

In the standard configuration, the tubes for removal of condensate are connected directly to the heat
exchangers. These protrude with the GL 25-12 tube connectors* (Borosilicate glass heat exchanger)
respectively with G 3/8“* thread joint (PVDF or stainless steel heat exchanger) above the base plate of
the cooler casing (Fig.2) (* standard).

Condensate removal is done by customer according to the type of operation with:

- Automatic float-type condensate traps AD-... only for over-pressure operation;

N O T E !

N O T E !

Stainless steel heat exchangers with G 3/8“ thread joint can be directly fitted
up with the float-type condensate trap AD-SS by means of a thread adapter
part number FF 11000

(1/2“ NPT to G 3/8“i). By this wall mounting of the AD-

SS unit isn’t necessary!


- Condensate collector container that is emptied manually;
- External peristaltic pump SR 25.1/Ex.