beautypg.com

Introduction, Description, Figure 1 – M&C TechGroup MP47_D Operator's manual User Manual

Page 7: Pump capacity mp47, 6introduction, 7description

background image

7

Gas sampling and gas conditioning technology

6-1.1.1-ME

6

INTRODUCTION

This diaphragm pump MP47 ... is suitable for corrosive gases. It is constructed especially for prob-
lems in the analysis technique.


7

DESCRIPTION

The pump MP47 has a PTFE-head. All sample contacting parts are out of PTFE. The pump works
absolutely lubricant free, so gases remain analytically unchanged. Due to a special diaphragm and
valve system, the pump operates maintenance-free.
The MP47 ... is available in 230 V or 115 V and can be equipped with an EEx motor for applications in
hazardous areas as well.
The standard MP47 ... is designed for normal applications at a maximum gas flow of 6 l/min. Up to
flow rates of 17 l/min we recommend the type MP47-Z.
With the option MP .../R a needle valve is build into the pump-head to adjust the flow rate (see data
sheet 6-1.1.3).
M&C provides for the series MP47 ... pump as option a diaphragm breakage monitoring. MP47 .../D is
equipped with a special double-membrane safety system (see 6-1.1.5).

Maximum temperature for medium during operating: -30 °C ... +80 °C.

The pump must be protected from the effectes of dust and water.

Before use please check the compatibility of the material.

Figure 1

Pump capacity MP47

0

50

100

150

200

250

300

350

400

450

0,4

0,5

0,6

0,7

0,8

0,9

1

1,2

1,4

1,6

1,8

2

2,2

l/hr

MP47

bar abs.

N O T E !


The diaphragm pump series MP47 ... are not allowed for liquid medium.