Gas sampling and gas conditioning technology

6-1.1.1-ME

Loosen the 4 screws M and remove the cover N ;

Screw the new diaphragm K into the rod L hand-tight;

Turn the flywheel P until the rod L is in a central position. Control that the bulge of the dia-
phragm K fits to the groove of housing A ;

Fix the diaphragm head B and the pressure plate C according to the mark;

Fix the 4 hexagon scews H constantly over cross until the plate springs R are flat;

Turn the flywheel P and check whether it works proper;

Mount the cover N ;


18.2

EXCHANGE OF THE VALVE PLATES

Loosen the 6 hexagon screws D and remove the pressure ring E ;

Change the o-ring S at the valve body F on the pressure side;

Change the valve plate T . Check whether the protection ring U is in the right position and the
6 layers of the valve plate T point to the top;

Screw in the valve bodies F and tighten them in a moderate way;

Change the o-ring S on the sucking side;

Remove the valve plate T and change it with a new one. Check whether the protection ring U
is in the right position and the 6 layers of the valve plate T point to the bottom;

Screw in the valve bodies G and tighten them in a moderate way;

Put on the pressure ring E and tighten the 6 hexagon screws.


18.3

CLEANING

When changing valve plates and diaphragm, inspect all parts for dirt before assembling the
pump head and clean them if necessary.

As far as possible clean the parts with a dry cloth. Solvents should not be used as they can
attack the plastics and synthetic rubber parts. If a compressed air line is available, blow the
parts out with it.

W A R N I N G !



Aggressive medium is possible.
Wear protective glasses and proper protective clothing during
disassembly, repair or cleaning!