TROUBLE SHOOTING

Before working on the pump isolate the power supply securely, then check that the lines are not live.
The following tips for fault-finding are best employed in the sequence shown.

Problem/Indication

Possible cause

Check/Action

Pump produces no
flow

No main supply

Check power supply; Check plug for correct
fit

Connections or lines are
blocked

Remove blockade

An external valve is closed or
a filter is blocked

Open valve or clean dirty/blocked filter

Liquid (condensate) has col-
lected in the pump head

Let the pump for a few minutes pumping air;

Flow, pressure or
vakuum too low

Diaphragm or valves are
weared out

Change diaphragm or/and valves

Compare the actual perfor
mance with the figures in
the technical data

The pump is not designed for this condition

There is pressure on the
pressure side and at the same
time vacuum or a pressure
above atmospheric on the
suction side

The pump is not designed for this condition

The

cross-section

of

the

pneumatic lines or connected
components is too small, or
they are restricted

To measure the performance, disconnect the
pump from the system (smaller diameter
tubing or a valve can significantly affect
performance)

A leak at a connector, in a
diaphragm/valve plate is
damaged or pump heads
dirty

Insulate the leak, tighten the screws, clean
or exchange dirty parts








N O T E !

If the pump does not operate properly and you cannot find any of the above
mentioned faults, send it back to M&C.
If you have been handling dangerous or highly aggressive gases please
clean the pump before dispatch.