Gas sampling and gas conditioning technology

5-2.1.5-ME

The following steps are recommended removing the filter housing lid (s. Fig. 2):

Turn to

ggle screw „A“ approx. 1 turn counter-clockwise, that the filter housing lid will lift up;

Turn handle „C“ in position „E“;

Turn clamp „B“ counter-clockwise;

Take toggle screw „A“ and pull out the filter housing lid;

Change filter element and possibly the lids;

Push in the filter housing lid into the filter area;

Turn clamp „B“ clockwise and turn with the handle „C“ the ringscrew „D“ in position „E“, that
the clamp „B“ will latch into the ringscrew „D“ and the threaded bolt „H“. It could be neccessary
to m

ove the ringscrew “D” on the thread rod of the toggle screw “A” into the correct position;

Turn handle „C“ in position „F“ and fasten the filter housing lid by turning the toggle screw „A“
clockwise by hand.























Figure 3 Dismounting of the filter housing lid













A

C, Pos. F

D

C, Pos. E

B

H