8
Gas sampling and gas conditioning technology
3-3.0-ME
8
TECHNICAL DATA
8.1
FOR BASIC COOLER ECM WITHOUT HEAT EXCHANGER
Gas cooler series EC
version ECM-1
version ECM-2
Part No.
02K7500 (a)**
02K7510 (a)**
Number of heat exchanger
possibility
prepared for 1 stream
prepared for 2 streams
Sample outlet dewpoint
range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C
Dew point stability
at const. conditions: < ±0,1°C
Sample inlet temperature
***max. +180°C
Sample inlet dew point
***max. +80°C
Gas flow rate
***max. 250l/h
***max. 2x150l/h
Ambient temperature
+10°C to +50°C
Storage temperature
-20 to +60°C
Total cooling capacity
40W at 10°C - 50°C ambient temperature
Ready for working
< 15 min
Power consumption
max. 200VA (start up current at 230V=2,5A; at 120V=4,5A)
Main power connection
240V/120V, 50Hz/60Hz (a):-15%/+10%
Electrical connections
terminals: 6-12mm , cable gland: 2xM20x1,5
Status alarm
1 free configurable status alarm with 2 potential free change-over
contacts, alarm point: < 2°C and > 8°C
contact rating 250VAC, 2A; 500 VA; 50W
Electrical safety
EN
EN 61010
CAN/CSA-C22.2 No.61010.1-04
UL Std. No. 61010-1(2 Edition)
System of protection
IP20 (EN 60529)
Case colour
RAL 9003
Method of mounting
wall mounting
Dimensions
Weight
270 x 270 x 316mm (WxHxD)
12kg/13,5kg (version: 230V/115V)
8.2
OPTIONS FOR BASIC COOLER ECM
Heat exchanger type
ECM-1G
ECM-1PV
ECM-1SS
ECM-2G
ECM-2PV
ECM-2SS
Part No
93K0140
93K0170
93K0160
97K0100
97K0110
97K0115
Material of heat
exchanger
Borosilicate
glass
PVDF
st. steel
316Ti
Borosilicate
glass
PVDF
st. steel
316Ti
Max. gas flow rate
250Nl/h
250Nl/h
250Nl/h
150Nl/h
150Nl/h
150Nl/h
Sample gas pressure
abs.
max. 2bar
(3bar)
1
max. 3bar
max. 10bar
max. 2bar
(3bar)
1
max. 3bar
max. 10bar
Sample gas connection
GL18
for Ø6mm
a.d. tube*
G1/4“i
G1/4“i
oder 1/4“
NPT**
GL18
for Ø6mm a.d.
tube*
Tube
Ø6mm*
Tube
Ø6mm*
Condensate connection
GL25 for
Ø12mm tube*
Ø8mm or
Ø10mm**
G3/8“i
G3/8“i o.
3/8“NPT**
GL25 for
Ø12mm tube*
Ø8mm or
Ø10mm**
G3/8“i
G3/8“i o.
3/8“NPT**
P at max. flow
1mbar
1mbar
1mbar
1mbar
1mbar
1mbar
Stagnant space approx.
100ml
100ml
100ml
40ml
25ml
30ml
Peristaltic pump SR25.2
1 pc. incorporated in the cooler, compl. installed, part no.: 01P9125; cooler weight
plus 0,6kg per pump
the sample gas pressure is limited to 2bar absolute
* Standard, other version on request
1
with GL connecting adapter
** Option
*** Maximum values in technical data must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew
point of 5 °C.