Gas sampling and gas conditioning technology

3-3.0-ME

16.2

CLEANING THE FINS OF THE CONDENSER

Dust on the fins of the condenser reduces the cooling capacity. Therefore it is necessary to clean the
fins from time to time. The following steps are recommended:

Shut off the gasflow;

Dismantle the tubing for gas in- and outlet;

Unscrew the cooler hood and remove it carefully;

Clean the fins carefully with compressed air;

Re-install the cooler hood;

Connect the tubing for gas in- and outlet.

N O T E !

Do not mix up the hose connections; the inlet and outlet connections of
the heat exchangers are marked with arrows.

16.3

MAINTENANCE OF THE OPTIONAL MOUNTED PERISTALTIC PUMP(S), TYPE
SR25.2

Before the maintenance work is carried out, it is necessary hat the specific safety procedures
pertaining to the system and operational process are observed !

D A N G E R


Dangerous voltage !
It is necessary to take the pump off the mains before any assembly,
maintenance and repair work is carried out !


Flexible tube, conveying belt, contact pulleys and contact springs are the only parts of the pump
subject to wear. They are simple to change.

16.3.1 CHANGE OF THE PUMP TUBE

C A R E !

Aggressive condensate is possible !
Wear protective glasses and proper protective clothing !

N O T E !

If you send back the peristaltic pump to the M&C service for repair,
please let us know what kind of condensate has been pumped.
Before sending the pump back clean all parts from dangerous or highly