beautypg.com

Technical data – M&C TechGroup SP 30-H.._EX2 Operator's manual User Manual

Page 10

background image

10

Gas sampling and gas conditioning technology

2-1.9.6EXME

10

TECHNICAL DATA

Serie SP

®

SP30-H1.1/EX2

SP30-H2/EX2

SP30-H1.1-V/EX2

Temperature sensor

Fe-CuNi

PT100 2-wire

Fe-CuNi

Temperature controller

external

Probe tube length L1

max. 1 m

Sample temperature

depending on the temperature class (see table temperature class). Please
specify with order.

Operating temperature

depending on the temperature class (see table temperature class). Please
specify with order.

Pre-filter

optional

optional

V20-2/30 insitu filter
length 520 mm,

60

mm, filter porosity 2

μm,

integrated and heated

Sample gas inlet
connection

G3/4"i DIN ISO 228/1

pre-filter with G1 1/2"i
DIN ISO 228/1

Dust loading

max. 2 g/m

3

>2 g/m

3

Probe tube volume

380 ml/m

420 ml

Sample pressure

max. 5 bar g

Ambient temperature

-20 °C to +70 °C

Storage temperature

-20 °C to +70 °C

Ready for operation

2 h

Power supply

230V 50/60Hz or 115V 50/60Hz

Heating capacity

0,6m: 600W, 1m: 800W, (other length/capacities on request)

Electrical connections

terminals, max. 4 mm

2

, 2x PG13,5 cable gland, terminal range 6-12mm

Electrical standard

EN 61010, EN60519-1

Degree of protection

IP54 EN 60529

Ex-marking

II 3 G Ex nA II T5 to T2 see table temperature class

Mounting flange

DN65 PN6, Form B with 4 welded screws on both sides M 12x 40 mm

Material of parts in contact
with the sample

stainless steel 1.4539

1.4539,
SS316Ti/SS316


Temperature class

Operating/sampling temperature
°C

Limiter
°C

T2

280

290

T3

185

190

T4

120

125

T5

85

90