Trouble shooting, Figure 8, Pma 10(s) with open housing – M&C TechGroup PMA 10S Operator's manual User Manual

Page 25

background image

25

Gas sampling and gas conditioning technology

9-1.1-ME

1 Quick-opening screw

5 Filter element

2 O

2

-transmitter-unit

6 Filter head with filter inlet

3 Filter body

7 Connection hose gas IN to filter

4 Filter element clamp


Figure 8

PMA 10(S) with open housing



21

TROUBLE SHOOTING

Error

Possible reason

Check/Repair


No indication


No supply voltage


Check supply voltage according to type plate.
Check wether mains cable is plugged in accu-
rate.
Check fine fuse in the low heat device socket.


No sample gas
flow


Sample line or filter is
blocked

Contamination of the in-
ternal diaphragm pump


Check sample system.


Loosen hose at the pump head and check
wether gas flow is present. Possibly open
pump head and clean valves, diaphragm and
pump head with a soap solution.