Measuring – M&C TechGroup PMA 10S Operator's manual User Manual
Page 23
23
Gas sampling and gas conditioning technology
9-1.1-ME
N O T E !
Always calibrate at the flow rate that is adjusted for the measurement
too.
Wait approximately 20
– 30 sec. until the indication has stabilised.
If necessary adjust span accurately according to the check gas concentration with a screw
driver at the span potentiometer in the front panel. In case of air e.g. to 20,9 % O
2
.
Check output signals at 20,9% O
2
:
Shut off pressure reducer valve and bottle- resp. Instrument air valve resp. integrated
sample gas pump.
Disconnect hose connection at the analyser.
Determination of the output signal:
(S
e
– S
np
) V resp. mA x gas concentration Vol % O
2
upper range value Vol % O
2
Se = Final value, signal output
Snp = Zero, signal output
Shut off pressure reducer output valve and bottle valve. Druckreglerausgangsventil und
Flaschenventil schließen. Disconnect hose connection at the analyser.
The span calibration is finished.
N O T E !
If during the span calibration great variations have to be co mpensated
(>2% O
2
) at the potentiometers, a second zero and span calibration is
reasonable.
18
MEASURING
For the first starting up at a new location, all steps in chapter 16 and 17 have to be performed.
By the requirements of precision the interval of the new calibration can be carried out daily or
weekly.
C A U T I O N !
The sample gas must be free from all liquid or solid particles, i.e. the dew
point of the gas must be below the equipment temperature so that no
condensate will occur inside the equipment. If necessary, lower the dew
point by means of a cooler or dryer. For dust filtration use a filter of 2
micron porosity !
We will be pleased to inform you about an optimal gas conditioning
The analyser now is ready for operation.
Output signal
Measurement
range 100 % O
2
Measurement
range 30 % O
2
0-1 V
0,209 V
0,697 V
0-20 mA
4,18 mA
13,93 mA
4-20 mA
7,34 mA
15,15 mA
+ S
np