ADJUSTMENT OF THE ALARM THRESHOLD

For adjustment of the alarm treshold on the front panel there is a potentiometer marked with
“alarm” and aside a push-button:

Actuate the push-button and keep it pressed. Now the digital indication of the indication

instrument shows the alarm threshold.

Adjust the alarm threshold in the range of 0

– 100 % O

2

at the alarm potentiometer with

suitable screw driver.

After adjustment of the desired alarm threshold release push button. The alarm threshold

now is adjusted and the digital indication again shows the current measurement.


The position of the measuring range selector switch is not relevant for the adjustment of the
alarm threshold.


15

CONNECTIONS ON THE REAR SIDE


Figure 7

Rear side

Low heat device socket with 2 fine fuses for power supply

Space for push-button for battery check - Option

Space for pump switch ON-OFF - Option

Space for 3-pole bushing for alarm contact output - Option