Zero calibration – M&C TechGroup PMA 10S Operator's manual User Manual
Page 19

19
Gas sampling and gas conditioning technology
9-1.1-ME
For span calibration with M&C O
2
-analysers it is possible to abandon on special test gas mixtures
because of the measuring principle and the linear measuring ranges. Dry and clean air is suffi-
cient. For measurement concentrations > 40% O
2
a calibration with corresponding test gas could
be possibly recommandable.
17.1
ZERO CALIBRATION
Connect a flexible PVC- or FEP-hose with the pressure reducer of the N
2
-zero-gas bottle.
The pressure reducer should have an output control range of max. 0
– 1,5 bar abs..
C A U T I O N !
The outlet pressure is only allowed to be adjusted at max. 0,1 bar. Other-
wise the measuring cell of the analyser will be destroyed.
Open the bottle valve and than the closed pressure reducer outlet valve and purge the
pressure reducer and the complete hose line for approximately 5 sec. the pressure reduc-
er and the complete hose line.
Check the adjusted control pressure and reduce if necessary to
≤ 0,1bar, then shut off the
pressure reducer valve again.
Connect the hose end of the zero-gas bottle connection to the gas inlet of the analyser.
Open the pressure reducer valve slowly, to avoid pressure peaks.
Adjust the flow rate to 50 Nl/hr at the flow meter.
N O T E !
Always calibrate at the flow rate that is adjusted for the measurement
too.
Wait approximately 20
– 30 sec. until the indication has stabilised.
If necessary adjust zero accurately to 0 % O
2
with a screw driver at the zero potentiometer
in the front panel
Check output signals at 0,0% O
2
:
N O T E !
If a gas mixture is analysed, the single gas components have to be checked
concerning potential cross sensitivity and regarded for zero calibration. (see
chapter 17.1.1 and 17.1.2).
Shut off pressure outlet valve and bottle valve.
Disconnect hose connection from the analyser. Schlauchverbindung am Analysator ent-
fernen.
Zero calibration is finished.
N O T E !
After zero calibration the span has to be calibrated too.
Output signal
Measurement
0-1 V
0 V
0-20 mA
0 mA
4-20 mA
4 mA