beautypg.com

Table 2-9: output menu -12, R to table 2-9 – Tektronix AFG3000 User Manual

Page 52

background image

Reference

2-12

AFG3000 Series Arbitrary/Function Generators Reference Manual

Output Menu

Table 2-9 shows the Output Menu. To access the Output menu, push the
front-panel Top Menu button, and then push the bottom bezel button.

Table 2-9: Output Menu

First Level

Second Level

Description

Load Impedance

50

Option buttons. Set the load impedance to 50

Ω.

Load

1

1. Range: 1

to 10 k

Ω.

When dBm is currently selected for output amplitude units, the amplitude units setting
is automatically changed to V

p-p

if you select high impedance.

Option buttons & Numeric input.
You can set the load impedance to any value
from 1

Ω to 10 kΩ.

High Z

Option buttons. Select high impedance.

Invert
Off On

Selects Invert On to invert a waveform.

Limit

High Limit

Numeric input.
Specifies the higher and lower amplitude limit.
See Level Meter on page 1-12.

Low Limit

Noise

Noise Add
Off On

Selects Noise Add On to add the internal noise
to a waveform.

Noise Level

2

2. Range: 0% to 50% of amplitude setting (V

p-p

) of signal waveform, resolution 1%

The maximum value of noise range is limited by waveform and amplitude range.

Sets the noise level.

External Add

3

Off On

3. AFG3100 and AFG3200 series only.

Selects External Add On to add an external
signal to the CH 1 output signal.