beautypg.com

Source[1|2]]:pulse:period, Source[1|2]]:pulse:transition[:leading – Tektronix AFG3000 User Manual

Page 143

background image

Syntax and Commands

AFG3000 Series Arbitrary/Function Generators Reference Manual

3-69

[SOURce[1|2]]:PULSe:PERiod

This command sets or queries the period for pulse waveform.

Group

Source

Syntax

[SOURce[1|2]]:PULSe:PERiod {|MINimum|MAXimum}

[SOURce[1|2]]:PULSe:PERiod?{MINimum|MAXimum}

Arguments

::=[]
where
is the pulse period.
::=[ns |

µs | ms | s]

Returns

Examples

SOURce1:PULSe:PERiod 200ns
sets the CH 1 pulse period to 200 ns.

[SOURce[1|2]]:PULSe:TRANsition[:LEADing]

This command sets or queries the leading edge time of pulse waveform.

Group

Source

Syntax

[SOURce[1|2]]:PULSe:TRANsition[:LEADing] {|MINimum|
MAXimum}

[SOURce[1|2]]:PULSe:TRANsition[:LEADing]?{MINimum|MAXimum}

Arguments

::=[]
where
is the leading edge time of pulse waveform.
::=[ns |

µs | ms | s]

Returns

Examples

SOURce1:PULSe:TRANsition:LEADing 200ns
sets the CH 1 leading edge time to 200 ns.