beautypg.com

Output[1|2]:impedance, Output[1|2]:impedance -32 – Tektronix AFG3000 User Manual

Page 106

background image

Syntax and Commands

3-32

AFG3000 Series Arbitrary/Function Generators Reference Manual

OUTPut[1|2]:IMPedance

The OUTPut:IMPedance command sets the output load impedance for the specified
channel. The specified value is used for amplitude, offset, and high/low level
settings. You can set the impedance to any value from 1

Ω to 10 kΩ. The default

value is 50

Ω .

The OUTPut:IMPedance? command returns the current load impedance setting in
ohms. If the load impedance is set to INFinity, the query command returns
“9.9E+37”.

Group

Output

Syntax

OUTPut[1|2]:IMPedance{|INFinity|MINimum|MAXimum}

OUTPut[1|2]:IMPedance?[MINimum|MAXimum]

Arguments

::=[]
where
::=OHM

INFinity sets the load impedance to >10 k

Ω.

MINimum sets the load impedance to 1

Ω.

MAXimum sets the load impedance to 10 k

Ω.

Returns

::=

Examples

OUTPut1:IMPedance MAXimum
sets the CH 1 load impedance to 10 k

Ω .