3) copy from other calibration channel, Enter, User cal ch id name ch copy ch reset – Konica Minolta Chroma Meter CS-200 User Manual
Page 72

70
M
ea
su
re
m
en
t
Pr
ep
ara
tio
n
(3) Copy from other calibration channel
7.
Press either
or
key to select
[CH COPY] and then
ENTER
key.
< COPY TO ??> screen appears.
8.
Press either
or
key to select
calibration channel to copy from.
To change calibration channel to copy to, press
SHIFT key to switch to SHIFT mode or press
or
key to select calibration channel to copy to.
Once copied, calibration channel to
copy to cannot be returned to previous
status before coping. Make sure to check
channel number before copying.
9.
Press
ENTER
key.
After “ PLEASE WAIT...” appears and the
value is copied,
When “ PLEASE WAIT...” appears, do not
switch OFF. Doing so may break stored data.