Error check – Konica Minolta Chroma Meter CS-200 User Manual
Page 115

113
De
scr
ipti
on
Error
No.
Symptom
Item to Check
Corrective Action
Page to
Refer
1
No display on LCD 
screen even after power 
is on.
Has AC adapter been 
properly plugged to AC 
outlet?
Connect AC adapter.
23
Has AC adapter been 
connected to this instru-
ment?
Connect AC adapter.
23
Is AC power source 
within rated?
Use within +/-10% of the 
nominal voltage.
8
Have batteries been 
placed?
Place batteries.
25
Haven't batteries 
drained?
Replace with new batter-
ies or connect with AC 
adapter.
25
2
Nothing is visible 
through finder.
Isn't lens cap on objec-
tive lens?
Remove lens cap.
-
Hasn't ND filter covered 
objective lens?
Do not use ND filter 
except for the case when 
object to measure is of 
high luminance.
-
Hasn't ND filter covered 
finder?
Do not use ND filter 
except for the case when 
object to measure is of 
high luminance.
-
3
Does not accept key 
operation.
Hasn't remote mode 
been set?
Press [ESC] key to can-
cel remote mode.
99
Don't you press unfunc-
tional key?
Press correct key.
-
Hasn't it been KEY 
LOCK?
Press [KEY LOCK] key 
for approx. 2 or more 
seconds to release KEY 
LOCK.
14
4
Measurement unavail-
able even after pressing 
measurement button
Doesn't menu screen 
appear?
Conduct measure-
ment with measurement 
screen appearing.
18
5
Gap between entered 
value in calibration 
value or target color 
and that appearing after 
setting
Calculation error could 
be found by 1 digit.
116
Error Check
Should error be found in this instrument, try corrective actions shown in the following 
table. If this does not help, this instrument has possibly been broken. Please contact the 
nearest KONICA MINOLTA SENSING authorized service facility with error number and 
version of your instrument. Version can be identified in procedure on p.116.