Opening and closing of finder shutter, Lv 92.74 – Konica Minolta Chroma Meter CS-200 User Manual
Page 48

46
S
e
tt
in
g
Opening and Closing of Finder Shutter
To prevent light from finder from badly influencing measurement, internal shutter of finder
is to close for every measurement. If observation through finder during measurement is
required, setting can be changed not to close finder shutter. In this case, light from finder
needs to be avoided by looking into finder during measurement.
∗ Setting at the shipment from factory : (automatically closes for every measurement)
Operation Procedure
1.
Press
ESC
key when menu or target
value setting menu appears.
Measurement screen appears on LCD screen.
2.
Make sure that [ ] appears on left
bottom indicating this instrument
has not been set for SHIFT mode.
If [SFT] appears instead, it means that
instrument is to operate in SHIFT mode.
Press
SHIFT key to cancel SHIFT mode.
2
3
1
S
U
H
U
T
T
E
R
CH00:
<
DEFAULT
>
Lv 92.74
cd/m
2
x 0.4185
y 0.4242
<
MODE
>
CONT.
[SFT ] [ 2° ]
<
ANGL
>
1° VIEW
<
SPD
>
AUTO
<
SYNC
>
NO SYNC
<
LENS
>
STANDARD
<
MEM
>
AUTO NUM
<
MEMORY DATA
>
M003:
Lv 1.04
x 0.3772
y 0.3663