5.
Press either
or
key
to switch update method of
measurement value directory.
If
key is pressed, mode switches in order of [AUTO
NUM] → [AUTOSAVE] → [MAN NUM] → [AUTO
NUM]. If kept pressed, this switches continuously.
In case of
key, mode switches in reversing
order. Keep pressing to switch continuously.
With [AUTO NUM] setting, memory channel number is replaced automatically
after storing data by pressing
MEMORY key. For example, if M005 is displayed,
measurement value is to store in M005 after
MEMORY key is pressed and number
of memory channel displayed in
If [AUTOSAVE] is set, measurement value will be saved automatically after
measurement and memory channel number changes accordingly. For example, if M005
is displayed, measurement value is to store in M005 after measurement is performed
and number of memory channel displayed in
In case of [MAN NUM], measurement value is to store memory channel of which number
is shown then. For example, if M005 is displayed, measurement value is to store in M005
and nunber of memory channel displayed in
With [AUTO NUM] or [AUTOSAVE], and M100 has been used up, data goes back to M000 again.
When data storage is to perform on memory channel with another
measurement value, a warning message appears.
Warning message “OK TO OVERWRITE?” appears. If OK, press
ENTER key, and if not, ESC key.
Setting is available not to display such message. See p.50.
Press
ESC key to stop.
6.
Press
ENTER
key.
Inversion cursor returns to
[MEMORY MODE]. Now change has been set.
7.
Press
ESC
key to return to
measurement screen.
remains even after switching OFF (O).
M
E
M
O
R
Y
M
O
D
E
CH00:
<
DEFAULT
>
Lv 20.80
cd/m
2
x 0.4476
y 0.4477
<
MODE
>
SINGLE
[ ] [ 2° ]
<
ANGL
>
1° VIEW
<
SPD
>
AUTO
<
SYNC
>
200.00Hz
<
LENS
>
STANDARD
<
MEM
>
AUTOSAVE
<
MEMORY DATA
>
M
OOO
:(NO DATA)
Lv ------
x ------
y ------
<
MENU
>
2/4
OBJECTIVE LENS STANDARD
DISPLAY DIGITS 4 FIGURES
MEMORY MODE
AUTO NUM
DATA PROTECT ON
BREAK : [ESC]
[ ] [ 2° ]
<
MENU
>
2/4
OBJECTIVE LENS STANDARD
DISPLAY DIGITS 4 FIGURES
MEMORY MODE
AUTOSAVE
DATA PROTECT ON
BREAK : [ESC]