Konica Minolta Chroma Meter CS-200 User Manual
Page 113

111
De
scr
ipti
on
Error message
Cause (Description)
Corrective Action
7
OVER
Luminance of measuring 
object is higher than avail-
able measurement range.
• Use ND filter and redo measure-
ment.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized 
service facility.
8
OFFSET ERROR
Zero calibration has not 
been correctly performed.
• Reset the power and redo meas-
urement.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized 
service facility.
9
EXCESSIVE LUMI-
NANCE VARIATION
Luminance change for 
measuring object is large.
• Place this instrument on stable
surface and redo measurement.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized 
service facility.
10
VIEWING-ANGLE 
SELECTOR ERROR
Measurement was per-
formed when measurement 
angle selector was in wrong 
position, or measurement 
angle selector was changed 
during measurement.
• Switch measurement angle selec-
tor and check that measurement 
angle appears in screen (------ is 
not displayed), and then redo 
measurement. In addition, do not 
operate measurement angle selec-
tor during measurement.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized 
service facility.
11
ROM ERROR
Data stored in ROM is 
broken.
• Do not switch OFF (O side) while
storing data or changing setting, or 
when message “PLEASE WAIT...” 
appears.
• If symptom does not improve,
please contact the nearest KONI-
CA MINOLTA SENSING authorized 
service facility.