Introduction, Note on use – Konica Minolta Chroma Meter CS-200 User Manual
Page 5

3
Introduction
This chroma meter realizes high-precision measurement of luminance and chromaticity
comparable to spectroradiometers by the employment of newly developed spectral fit-
ting method. Carefully read this manual before using one.
Packaging material
Retain accompanying packaging materials (carton, protector, and plastic bag) and
holding cap (CS-A24) supplied as standard accessory for future usage.
This is delicate measurement instrument. Use packaging materials supplied in pur-
chasing in case this instrument needs to be transferred for such purpose as main-
tenance in KONICA MINOLTA SENSING's factories. These packaging materials are
useful for minimizing shock or vibration to this instrument in such situation. Use hold-
ing cap for the same purpose especially to protect optical system of this instrument.
Should any of these packaging materials or holding cap (CS-A24) be lost or broken,
please contact the nearest KONICA MINOLTA SENSING authorized service facility.
Note on Use
Operating Environment
• Do not use this instrument outdoor since standard accessory AC adapter is designed
for indoor use.
• Do not disassemble this instrument for being composed of delicate electronic compo-
nents.
• Use this instrument at rated voltage of 100 V - 120 V
or 200 V - 240 V
.
Connect AC power cord to AC outlet with rated voltage and frequency. Connected
voltage should not be outside the range of +/-10% of nominal.
• This instrument is classified into a Pollution Degree 2 as instrument used in mainly
in manufacturing plant, laboratory, warehouse or equivalents. Use this instrument in
metal dust free and non condensing potential environment.
• This instrument is categorized into Installation Category II as equipment connected to
commercially available power source.
• Connect PC for controlling this instrument to the outlet with protective grounding.
Failure to follow this may result in electric shock due to short circuit.
• Take care not to enter foreign substance like water or metal in this instrument. Operat-
ing in such state cause serious danger.
• Do not use this instrument under direct sunlight or near heater. The internal tempera-
ture of this instrument to becomes much higher than ambient temperature which may
break this instrument.
• Avoid rapid change in ambient temperature which may form dew condensation.
• Avoid using this instrument in extremely dusty or humid place.
• Use this instrument at ambient temperature between 0 and 40˚C and relative humidity
85% or less (at 35˚C) with no condensation. Operating this instrument outside speci-
fied temperature and humidity range may unsatisfy its original performance.