Scan:project, Scan:project:create, Scan:project:currently:cycle – RIGOL M300 User Manual
Page 64: Scan:task:add

RIGOL
Programming Guide for DM3000
2-52
1. :SCAN:PROJect?
Syntax
:SCAN:PROJect?
Function
Query the name of the scan project currently created. When there is
no scan project, the query returns NULL.
2. :SCAN:PROJect:CREAte
Syntax
:SCAN:PROJect:CREAte?
:SCAN:PROJect:CREAte
Function
Create a scan project with the specified name.
Query whether a task has been created for the current scan project.
The query returns True if an available task exists; otherwise, returns
False.
Explanation
should be composed of letters a - z, A - Z and numbers
within 0 and 9 and its length can not exceed 15 characters.
3. :SCAN:PROJect:CURRently:CYCLe?
Syntax
:SCAN:PROJect:CURRently:CYCLe?
Function
Query the number of cycles of the current scan project.
The query returns an integer.
Explanation Refer to :SCAN:RUN to set the number of cycles and run the scan
project.
4. :SCAN:TASK:ADD
Syntax
:SCAN:TASK:ADD{
Function
Add a task to the current scan project.
Explanation
If the tasks previous to this task are not added, the multimeter
will automatically add these tasks using the current
configurations.
range is from 1 to 16. Wherein, for DCV, ACV, 2WR, FREQ, PERI,
CAP and DIODE, channel 1 to channel 12 can be used; while, for
DCI and ACI, channel 13 to channel 16 can be used.
task and it can be
DCV|ACV|DCI|ACI|RESistance|DIODe|CAPacitance|PERIod|FRE