Trigger:auto:hold:sensitivity, Trigger:single, Trigger:single:trigger – RIGOL M300 User Manual
Page 49: Trigger:ext

RIGOL
Programming Guide for DM3000
2-37
:TRIGger:AUTO:HOLD {ON|OFF|1|0}
Function
Enable or disable the auto trigger delay function.
The query returns ON or OFF.
Default
OFF
4. :TRIGger:AUTO:HOLD:SENSitivity
Syntax
:TRIGger:AUTO:HOLD:SENSitivity?
:TRIGger:AUTO:HOLD:SENSitivity {0|1|2|3|MIN|MAX|DEF}
Function
Set the sensitivity of the auto trigger delay.
The query returns 0, 1, 2 or 3.
Explanation
The sensitivity corresponding to each parameter is as follows.
Value
Sensitivity
0
0.01%
1
0.1%
2
1%
3
10%
MAX
10%
MIN
0.01%
DEF
0.1%
5. :TRIGger:SINGle
Syntax
:TRIGger:SINGle?
:TRIGger:SINGle {
|MIN|MAX|DEF}
Function
Set the number of samples for single trigger.
Explanation
ranges from 1 to 1000.
The “DEF” is 1.
6. :TRIGger:SINGle:TRIGger
Syntax
:TRIGger:SINGle:TRIGger
Function
Execute single trigger.
Explanation
First, select the single trigger source using the :TRIGger:SOURce
SINGLE command; otherwise, this command is invalid.
7. :TRIGger:EXT
Syntax
:TRIGger:EXT?
:TRIGger:EXT {RISE|FALL}
Function
Set the external trigger type to RISE or FALL.