Chapter 2 Command System

Programming Guide for DM3000

2-48

Syntax

:DATAlog:CONFigure:STARtmode:EXTern

Function

Set the start mode of the current Datalog function to External.

Explanation The instrument enters wait-for-trigger state after recieving this

command and starts the data acquisition when trigger signal is
recieved.


6. :DATAlog:CONFigure:STARtmode:DELAytime
Syntax

:DATAlog:CONFigure:STARtmode:DELAytime?
:DATAlog:CONFigure:STARtmode:DELAytime

Function

Set the delay time for Auto Datalog and the unit is s.
The query returns an integer.

Explanation

can be any integer ranging from 0 to 3600.

7. :DATAlog:CONFigure:STOPmode:TIME
Syntax

:DATAlog:CONFigure:STOPmode:TIME?
:DATAlog:CONFigure:STOPmode:TIME

Function

Set the duration of data acquisition. The data acquisition stops
automatically when the specified time is reached.
The query returns an integer.

Explanation The range of

is relevant with the sample rate and the largest

range is from 1 to 2097150 and the unit is s.


8. :DATAlog:CONFigure:STOPmode:NUMber
Syntax

:DATAlog:CONFigure:STOPmode:NUMber?
:DATAlog:CONFigure:STOPmode:NUMber

Function

Set the number of data acquisitions. The data acquisition stops
automatically when the specified sample number is reached.
The query returns an integer.

Explanation

can be any integer ranging from 1 to 2097151.


9. :DATAlog:CONFigure:RATE
Syntax

:DATAlog:CONFigure:RATE?
:DATAlog:CONFigure:RATE

Function

Set the sample rate of data acquisition.

Explanation

ranges from 1 to 13 and the sample rate and return value

corresponding to each range are as follows: