Resolution:current:dc – RIGOL M300 User Manual
Page 35

RIGOL
Programming Guide for DM3000
2-23
:RESOlution:VOLTage:AC {0|1|2|MIN|MAX|DEF}
Function
Set the reading resolution of AC voltage measurement.
The query returns 0, 1 or 2.
Explanation
The AC voltage measurement function must be enabled before
using this command.
The “DEF” is 1.
The reading resolution corresponding to each parameter is as
follows.
Range
Reading resolution
0
3 ½ digits
1
4 ½ digits
2
5 ½ digits
MAX
5 ½ digits
MIN
3 ½ digits
DEF
4 ½ digits
Example
Set the reading resolution of AC voltage measurement to 5 ½
digits:
:RESOlution:VOLTage:AC 2
4. :RESOlution:CURRent:DC
Syntax
:RESOlution:CURRent:DC?
:RESOlution:CURRent:DC {0|1|2|MIN|MAX|DEF}
Function
Set the reading resolution of DC current measurement.
The query returns 0, 1 or 2.
Explanation
The DC current measurement function must be enabled before
using this command.
The reading resolution of each parameter.
Range
Reading resolution
DM306x
DM305x
0
4 ½ digits
3 ½ digits
1
5 ½ digits
4 ½ digits
2
6 ½ digits
5 ½ digits
MAX
6 ½ digits
5 ½ digits
MIN
4 ½ digits
3 ½ digits
DEF
5 ½ digits
4 ½ digits
Example
Set the reading resolution of DC current measurement to 5 ½
digits: