Measure:current:ac:frequency:state, Measure:resistance, Measure:resistance:range – RIGOL M300 User Manual
Page 27

RIGOL
Programming Guide for DM3000
2-15
secondary display is enabled in AC current measurement.
30. :MEASure:CURRent:AC:FREQuency:STATe?
Syntax
:MEASure:CURRent:AC:FREQuency:STATe?
Function
Query whether the frequency is displayed on the secondary screen
during AC current measurement.
The query returns DISPLAY or HIDE.
31. :MEASure:RESistance?
Syntax
:MEASure:RESistance?
Function
The query returns the 2-wire resistance measurement value in
scientific notation and the unit is Ω.
32. :MEASure:RESistance
Syntax
:MEASure:RESistance {0|1|2|3|4|5|6|MIN|MAX|DEF}
Function
Set the range of 2-wire resistance measurement.
Explanation The measurement mode will change to “Manual” while you set the
range.
The range corresponding to each parameter is as follows.
Parameter
Range
DM306x
DM305x
0
200 Ω
400 Ω
1
2 kΩ
4 kΩ
2
20 kΩ
40 kΩ
3
200 kΩ
400 kΩ
4
1 MΩ
4 MΩ
5
10 MΩ
100 MΩ
6
100 MΩ
MAX
100 MΩ
100 MΩ
MIN
200 Ω
400 Ω
DEF
200 kΩ
400 kΩ
Example
Set the range of 2-wire resistance measurement to the minimum:
:MEASure:RESistance 0 or
:MEASure:RESistance MIN
33. :MEASure:RESistance:RANGe?
Syntax
:MEASure:RESistance:RANGe?