Measure:resistance:digit, Measure:fresistance, Measure:fresistance:range – RIGOL M300 User Manual
Page 28: Measure:fresistance:digit

RIGOL
Programming Guide for DM3000
2-16
Function
Query the current range of 2-wire resistance measurement.
The query returns 0, 1, 2, 3, 4, 5 or 6.
34. :MEASure:RESistance:DIGIt
Syntax
:MEASure:RESistance:DIGIt?
:MEASure:RESistance:DIGIt {INC|DEC|5|6|7}
Function
Set the display digit of 2-wire resistance measurement value.
The query returns 5, 6 or 7.
Explanation Refer to the “Explanation” in :MEASure:VOLTage:DC:DIGIt.
Example
Set the display digit of 2-wire resistance measurement value to 6:
:MEASure:RESistance:DIGIt 6
Decrease the display digit by one:
:MEASure:RESistance:DIGIt DEC
35. :MEASure:FRESistance?
Syntax
:MEASure:FRESistance?
Function
The query returns the 4-wire resistance measurement value in
scientific notation (for example, +2.366031E+03) and the unit is Ω.
36. :MEASure:FRESistance
Syntax
:MEASure:FRESistance {0|1|2|3|4|5|6|MIN|MAX|DEF}
Function
Set the range of 4-wire resistance measurement.
Explanation Refer to the “Explanation” in :MEASure:RESistance.
Example
Set the range of 4-wire resistance measurement to the minimum:
:MEASure:FRESistance 0 or
:MEASure:FRESistance MIN
37. :MEASure:FRESistance:RANGe?
Syntax
:MEASure:FRESistance:RANGe?
Function
Query the range of 4-wire resistance measurement.
The query returns 0, 1, 2, 3, 4, 5 or 6.
38. :MEASure:FRESistance:DIGIt
Syntax
:MEASure:FRESistance:DIGIt?
:MEASure:FRESistance:DIGIt {INC|DEC|5|6|7}
Function
Set the display digit of 4-wire resistance measurement value.
The query returns 5, 6 or 7.