1. :MEASure?
Syntax
:MEASure?
Function
Query whether the current measurement is finished. If yes, the query
returns TRUE; otherwise, the query returns FALSE.
2. :MEASure
Syntax
:MEASure {AUTO|MANU}
Function
Set the measurement mode to Auto or Manual.
Default
AUTO
3. :MEASure:VOLTage:DC?
Syntax
:MEASure:VOLTage:DC?
Function
The query returns the measurement value of DC voltage in scientific
notation (for example, +2.53021747E-04) and the unit is V.
4. :MEASure:VOLTage:DC
Syntax
:MEASure:VOLTage:DC {0|1|2|3|4|MIN|MAX|DEF}
Function
Set the range of DC voltage measurement.
Explanation The measurement mode will change to “Manual” while you set the
range.
The range corresponding to each parameter is as follows.
Parameter
Range
DM306x
DM305x
0
200 mV
400 mV
1
2 V
4 V
2
20 V
40 V
3
200 V
400 V
4
1000 V
1000 V
MIN
200 mV
400 mV
MAX
1000 V
1000 V
DEF
20 V
40 V
Example
Set the range to the minimum:
:MEASure:VOLTage:DC 0 or
:MEASure:VOLTage:DC MIN
5. :MEASure:VOLTage:DC:RANGe?
Syntax
:MEASure:VOLTage:DC:RANGe?