Calculate:limit:lower, Calculate:limit:upper – RIGOL M300 User Manual
Page 56

RIGOL
Programming Guide for DM3000
2-44
The query returns PASS or FAIL.
Explanation Limit operation function must be enabled before sending this
command.
17. :CALCulate:LIMIt:LOWEr
Syntax
:CALCulate:LIMIt:LOWEr?
:CALCulate:LIMIt:LOWEr {
|MIN|DEF}
Function
Set the lower limit of Limit operation.
The query returns the lower limit of Limit operation under the current
measurement function in scientific notation.
Explanation The lower limit should not be greater than the upper limit. For
more details, refer to :CALCulate:LIMIt:UPPEr.
The range of
depends on the current measurement
function.
Measurement
Default
Unit
DC voltage
±1200
0
V
AC voltage
0 - 900
0
V
DC current
±12
0
A
AC current
0 - 12
0
A
Resistance
0 - 1.2e+08
0
Ω
Capacitance
0 - 2.4e-04
0
F
Frequency
3 - 3.0e+05
3
Hz
Period
3.0e-06 - 3.0e-01
3.0e-06
s
Ratio
±1.0e+09
-1.0e+09
18. :CALCulate:LIMIt:UPPEr
Syntax
:CALCulate:LIMIt:UPPEr?
:CALCulate:LIMIt:UPPEr {
|MAX|DEF}
Function
Set the upper limit of Limit operation.
The query returns the upper limit of Limit operation under the current
measurement function in scientific notation.
Explanation The upper limit should not be lower than the lower limit. For more
details, refer to :CALCulate:LIMIt:LOWEr.
The range of
depends on the current measurement
function.
Measurement
Range
Default
Unit
DC voltage
±1200
1
V