Programming Guide for DM3000
2-12
17. :MEASure:VOLTage:AC:FREQuency:HIDE
Syntax
:MEASure:VOLTage:AC:FREQuency:HIDE
Function
Hide the frequency on the secondary screen during AC voltage
measurement.
Explanation The command is valid only when the frequency display on the
secondary display is enabled in AC voltage measurement.
18. :MEASure:VOLTage:AC:FREQuency:STATe?
Syntax
:MEASure:VOLTage:AC:FREQuency:STATe?
Function
Query whether the frequency is displayed on the secondary screen
during AC voltage measurement.
The query returns DISPLAY or HIDE.
19. :MEASure:CURRent:DC?
Syntax
:MEASure:CURRent:DC?
Function
The query returns the DC current measurement value in scientific
notation (for example, -3.74725404E-06) and the unit is A.
20. :MEASure:CURRent:DC
Syntax
:MEASure:CURRent:DC {0|1|2|3|4|MIN|MAX|DEF}
Function
Set the range of DC current measurement.
Explanation The measurement mode will change to “Manual” while you set the
range.
The range corresponding to each parameter is as follows.
Parameter
Range
0
2 mA
1
20 mA
2
200 mA
3
1 A
4
10 A
MIN
2 mA
MAX
10 A
DEF
200 mA
Example
Set the range of DC current measurement to the maximum:
:MEASure:CURRent:DC 4 or
:MEASure:CURRent:DC MAX