User’s Guide for DM3058/DM3058E

2-53

 Press

two times until this button is glittery, then a trigger will be hold and

a stable reading will be captured and hold on the screen.

Press

 Auto  Hold, the display shows:

Figure 2-68 Manu interface of Reading Hold

Table 2-23 Reading Hold Function Menu Description

Function Menu

Settings

Description

On/Off

Turn on/off the reading hold function.

0.01

Set the hold scope as 0.01%.

0.1

Set the hold scope as 0.1%.

1%

Set the hold scope as 1%.

10%

Set the hold scope as 10%.

Save all changes, back to a higher level

menu.


Reading Hold Function
Start the Reading Hold Function, the hold measurement use the following rules
judge the reading count:

When Max () - Min () ≤ hold range x Readings, the multimeter hold Reading on the
display.
The display update a new reading basing on the current value and the three
preceding measurements before the reading was hold:
Max (Reading, ReadingN-1, ReadingN-2, ReadingN-3)
Min (Reading, ReadingN-1, ReadingN-2, ReadingN-3)

NOTE:
 The default value of hold range is 0.1%, and the range setting is

stored in volatile memory.

 When Reading Hold started, the input resistance was set to 10MΩ for

all DC voltage ranges to reduce noise arising from the open-loop

testing.