background image

RIGOL

User’s Guide for DM3058/DM3058E

2-26

Table 2-10 Frequency Test Characteristics

Ranges

200mV, 2V, 20V, 200V, 750V

Input Signal Range

20Hz ~ 1MHz

Input Protection

750V

RMS

on all ranges (HI

terminal)

Configurable Parameters

REL


3. Set the relative value (Optional).

Press REL to open or close Relative operation function. When it is open, “REL”
is shown over the screen and the reading displayed is a relative value which
comes from the result of Actual measurement value subtracts the value that
has been set. (For the details about how to set, please refer to

Math

Functions in Chapter 2.)

4. Read measurement value

Frequency measurement is fixed at “Slow” rate, 5.5 digits resolution. Therefore,
you cannot adjust the reading rate when reading the result.

5. View history data

Press History to enter into the following interface and check the data getting
from this measurement and saved.

Figure 2-32 View interface of History information


There are three types of way for viewing historical date: “Info”, “List” and
“HistoG”. Press Save after checking. By pressing Update, the History
information will be refreshed to be current newest information.