RIGOL

User’s Guide for DM3058/DM3058E

6-3


DC Voltage
Input Resistance:

200mV and 2V range 10MΩ or >10GΩ selectable
(Input signals which exceed ±2.5V in these ranges will pass the 100kΩ
(typical) clamp resistance.
20V, 200V and 1000V range 10MΩ±2%

Input offset current:

<90pA, 25℃

Input Protection:

1000V on all ranges

CMRR (common mode rejection ratio): 120dB (For the 1kΩ unbalanced resistance in LO lead,

maximum ±500VDC).

NMRR (normal mode rejection ratio): 60dB at “slow” measurement rate

Add 20dB when open the “AC filter” (Settling time adds
0.35s (source impedance near zero))


Resistance
Measurement Method:

Selectable 4-wire or 2-wire.

Current source referenced to LO input.

Open-circuit Voltage:

Limit in <8V.

Max. Lead Resistance:

10% of range per lead for 200Ω, 1kΩ per lead (4-wire ohms) on all
other ranges.

Input Protection:

1000V on all ranges.


DC Current
Shunt Resistor:

200μA sampling voltage<8mV
2mA sampling voltage<80mV
1Ω for 20mA, 200mA
0.01Ω for 2A, 10A

Input Protection:

Externally accessible 10A, 250V fast-melt fuse

Internal 12A, 250V slow-melt fuse

Continuity / Diode Test
Measurement Method:

1mA ±5%

current source

, <8V

open-circuit voltage

Response Time:

123 samples / sec, with beeper

Continuity Threshold:

Adjustable from 1Ω to 2000Ω

Input Protection:

1000V


Setting time attentions:
Settling time in voltage measurement is affected by the source impedance, media characteristics
of the cable and input signals.