beautypg.com

Loading errors (ac volts), Loading errors (ac volts) -4 – RIGOL DM3058 User Manual

Page 122

background image

RIGOL

User’s Guide for DM3058/DM3058E

5-4

Loading Errors (AC Volts)

In the AC Voltage function, the input of DM3058/DM3058E appears as a 1MΩ
resistance in parallel with 100 pF of capacitance. The test lead that you use to
connect signals to the multimeter will also add additional capacitance and loading.
Table 5-2 shows the multimeter’s approximate input resistance at various
frequencies.

Table 5-2 Approximate input resistance at various frequencies

Input Frequency

Input Resistance

100Hz

1MΩ

1kHz

850kΩ

10kHz

160kΩ

100kHz

16kΩ


For low frequencies:

Error (%)

%

100

1M

Rs

R

-

s

×

+

=


Additional error for high frequencies:

Error (%)

%

100

1

)

2

(

1

1

Ч

Ч

Ч

Ч

+

=

in

s

C

R

F

π

s

R

: source resistance

F

: input frequency

in

C

: input capacitance (100 pF) plus test lead capacitance


This manual is related to the following products: