User’s Guide for DM3058/DM3058E

2-52

Auto Triggering


“Interval” parameter needs to be set in auto trigger mode. Press

on the front

panel to enable the auto trigger and acquire continue readings. Furthermore, auto
trigger is the default mode at digital multimeter power on.

Press

 Auto, the display shows:




Figure 2-67 Setting interface of Auto Trigger

Table 2-22 Auto Trigger Function Menu Description


1. Interval
 The interval means the waiting time before sampling since trigger signal has

been sent. The interval is different under different measure speed.
Fast: the default interval is 8ms, setting range is 8ms~2000ms.
Middle: the default interval is 50ms, setting range is 50ms~2000ms.
Slow: the default interval is 400ms, setting range is 400ms~2000ms.

The default measure speed is slow, so the default interval is 400ms.

 The interval is stored in volatile memory.

2. Reading Hold
 The reading hold mode captures and holds a reading on the front panel display

when a stable reading is detected. The system Hold ranges are 0.01%, 0.1%,
1%, and 10%.


Function Menu

Settings

Description

Interval

Set interval time in 400~2000ms.

Hold

ON/OFF

Turn the Reading Hold function ON or

OFF.

Save all changes, back to a higher level

menu.