Auto triggering, Auto triggering -52 – RIGOL DM3058 User Manual
Page 74
RIGOL
User’s Guide for DM3058/DM3058E
2-52
Auto Triggering
“Interval” parameter needs to be set in auto trigger mode. Press
on the front
panel to enable the auto trigger and acquire continue readings. Furthermore, auto
trigger is the default mode at digital multimeter power on.
Press
Auto, the display shows:
Figure 2-67 Setting interface of Auto Trigger
Table 2-22 Auto Trigger Function Menu Description
1. Interval
The interval means the waiting time before sampling since trigger signal has
been sent. The interval is different under different measure speed.
Fast: the default interval is 8ms, setting range is 8ms~2000ms.
Middle: the default interval is 50ms, setting range is 50ms~2000ms.
Slow: the default interval is 400ms, setting range is 400ms~2000ms.
The default measure speed is slow, so the default interval is 400ms.
The interval is stored in volatile memory.
2. Reading Hold
The reading hold mode captures and holds a reading on the front panel display
when a stable reading is detected. The system Hold ranges are 0.01%, 0.1%,
1%, and 10%.
Function Menu
Settings
Description
Interval
Set interval time in 400~2000ms.
Hold
ON/OFF
Turn the Reading Hold function ON or
OFF.
Save all changes, back to a higher level
menu.