Capacitance characteristics, Capacitance characteristics -8 – RIGOL DM3058 User Manual
Page 132
RIGOL
User’s Guide for DM3058/DM3058E
6-8
Capacitance Characteristics
Accuracy Specifications (% of reading + % of range)
[1,2]
Function
Range
[2]
Test Current
1 Year
23℃±5℃
Temperature
Coefficient
0 ℃ to 18 ℃
28 ℃ to 55 ℃
Capacitance 2.000nF
200nA
3 + 1.0
0.08+0.002
20.00nF
200nA
1 + 0.5
0.02+0.001
200.0nF
2μA
1 + 0.5
0.02+0.001
2.000μF
10μA
1 + 0.5
0.02+0.001
200μF
100μA
1 + 0.5
0.02+0.001
10000μF
1mA
2 + 0.5
0.02+0.001
NOTE:
[1] Specifications are for 30 minute warm–up using REL operation. Additional errors may occur
for non–film capacitors.
[2] Specifications are for 1% to 110% of range on 2 nF range and 10% to 110% of range on all
other ranges.
Measurement Characteristics
Capacitance
Measurement Method:
Current input with measurement of resulting ramp.
Connection Type:
2-wire
Input Protection:
1000V on all ranges
Measurement attentions:
Measurement of small capacitance is easily affected by external noise thus to cause measurement
error, disable input will reduce this error.