Example 6: thermocouple setting and measurement, Example 6: thermocouple setting and measurement -7, Example 6 – RIGOL DM3058 User Manual
Page 105: Thermocouple setting and measurement
![background image](https://www.manualsdir.com/files/859132/content/doc105.png)
RIGOL
User’s Guide for DM3058/DM3058E
3-7
Example 6:
Thermocouple Setting and Measurement
Thermocouple is a common temperature sensor. Type, voltage and cold-side
temperature are available for thermocouple measurement.
The instrument can measure temperature of HI and LO terminals (cold-side), and
calculate absolute temperature of hot-side. Besides, only the relationship between
thermocouple voltage and cold-side temperature is need to be input according
thermocouple type when you set the thermocouple sensor.
Operating Steps:
About the connection of Sensor please refer to “To Measure Sensor” in chapter 2.
Figure 3-12 Sketch map for connecting
1. Press
New to enter into the setting interface.
Figure 3-13 New interface
2. Press Prpty to enter into the edit interface of Property.
Figure 3-14 Edit interface