Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©

Page 82

www.PressureSystems.com

scanners within such a larger system, a simple three part addressing scheme
is available. Individual scanners are identified by their Cluster, Rack, and
Slot (CRS) address, a three-digit hexadecimal numbe

r.

A cluster (the most significant address unit), is numbered 0-F (0-15 decimal)
for a total of 16 clusters (clusters often refer to other geographical units like
control rooms or test stands).

A

rack (the middle address digit), is numbered 0-F (0-15 decimal) for a total

of 16 racks per cluster. (NOTE: The 98RK-1 front-panel switch for setting the

rack address is labeled 0-15 decimal instead of hex 0-F)


A

slot (the least significant address digit), numbered 1 through 8, designates

a particular 9816 pressure scanner in a rack.


Example:

Request model number from any NetScanner

System module:


q00


Read response indicating a Model 9816:


"9816"


Request TCP back-off delay for any NetScanner

System module:


q07


Read hex (16-bit binary) response:


"001F" (31 decimal, or 31*20= 620 µsec.)