Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©

Page 23

www.PressureSystems.com




















Figure 2.5


Each 9816 is manufactured with a connection to one of the eight (8) CAL port ranges, based on
the unit’s full scale pressure range. Assignment of 9816's CAL range is show in Table 2.1 (next
page). In the case of a 9816 with multiple pressure ranges installed, the unit will use the CAL
range of the lowest installed full scale pressure range.

Note

CAL port assignments are based on a pressure range, NOT 98RK-1
slot assignments.

Table 2.1

Pneumatic Connector Assignment

98RK-1

Connection

Pressure Scanner Range Input

Cal 1

User defined

Cal 2

10" W.C.; 20" W.C.; ±1 psid, ±1.5 psid

Cal 3

±2.5; ±5 psid

Cal 4

±10; ±15 psid, ±20 psid, ±25 psid

Cal 5

±30; ±45; ±50 psid

Cal 6

±75 psid, ±100 psid

Cal 7

+150; 200; 250; 300 psid

Cal 8

500 psid, 600 psid, 650 psid, 750 psid, 850
psid

This manual is related to the following products: