Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©

Page 135

www.PressureSystems.com





























Figure 5.12

Update Firmware Screen

Click the “Download” button to begin sending the file to the module. A progress

bar displays how much of the file has been sent. This “Status” field will display

the status of the download.


The module automatically “re-boots" for acceptance of the new firmware. This

entire process requires approximately two (2) minutes.

WARNING: While updating module firmware, DO NOT power-cycle your 98RK-1,

the 9816 scanner, or your PC. If the firmware update procedure is

interrupted by any of these practices, the module may be left in a

“permanent” inoperable state with no operable firmware to reboot it.

The ONLY acceptable way to interrupt the firmware update process is

to select “Abort Download before the flash memory begins to be

overwritten. In the event of such failure, module operation can only be

restored by unplugging the ‘bad” flash chip and installing a “good”

flash chip containing a valid working firmware version.”