Example:

● Finish

the

Multi-Point Calibration function previously started (as indicated by the

previous examples of 'C' sub-commands ‘00’ and ‘01’). Calculate new adjustment

coefficients, and save them in the non-volatile memory of the module’s transducers.

These new coefficients will then be used for all subsequently calculated EU data

acquired by the module, until another calibration function is performed in the future.


C 02

w08

w09


Read responses (separately for each command executed above):

"A"

"A"

"A"


If an error (“N”) response is returned on the first command, either the correct number of

calibration points (per ‘00’ sub-command) were not supplied with reasonable pressure

data values (via the multiple ‘01’ sub-commands), or the collected data yielded new

calculated coefficients with unreasonable values.


If execution of the last two 'w' commands is skipped above, the new calibration data

obtained will be stored only in volatile storage, and will be available for use only until the

module is RESET or loses power.