Pressure Systems, Inc. 98RK-1 & 9816 User’s Manual©

Page 55

www.PressureSystems.com

Note

The calculated gain values from the latest 'Z' command will be
lost when the module is powered OFF. To save these gain

terms to each transducer's nonvolatile memory, refer to the Set
Operating Options
('w') command (index 09).


Example:

Send TCP/IP command to a 9816 module (via its open socket) to calculate and set gain

coefficients for channels 8 through 4. Instruct the module to use 14.8890 psi as the

applied pressure instead of each transducer’s full-scale value:


Z00F8 14.8890

Response:

" 1.01289 1.06953 1.03750 0.99704 1.10051"

Actual gain values are returned in the above response as decimal ASCII strings, each

preceded by a space character. From left-to-right: they are for channels 8, 7, 6, 5, and

4.