System Range Codes
The following range codes are stored in each DH200 and digital 9400 and 9401 pressure
transducer. The range code of each transducer can be read through the Read Internal
Coefficient (‘u’) command, array index 01-10 hex, coefficient index 0A. Standard 9816 Range
Codes are Bold.
Range Code
Full Scale Pressure
Minimum Calibration
Pressure
1
±.360 psi (±10" Water
Column)
-0.360 psi
2
±.720 psi (±20" Water
Column)
-0.720 psi
3
±1 psid
-1.0 psi
4
±2.5 psid
-2.5 psi
5
±5 psid
-5 psi
6
10 psid
-5 psi
7
15 psid
-5 psi
8
30 psid
-5 psi
9
45 psi
0 psi
10
100 psi
0 psi
11
250 psi
0 psi
12
500 psi
0 psi
13
600 psi
0 psi
14
300 psi
0 psi
15
750 psi
0 psi
16
10 psid
-10 psi
17
15 psid
-12 psi
18
30 psid
-12 psi
19
45 psid
-12 psi